ND-1000S/2000 Waterproof Micro Sievert Survey Meter
Waterproof micro sievert survey meter 3 Range, 0-1000 Sv/hr
Features:
- Detects Gamma and X-ray.
- Rugged, waterproof, dustproof, nylon reinforced fiberglass case.
- Drop tested 1 meter (onto concrete) in excess of 20 times with no damage or failures.
- Battery/high voltage test position on range switch.
- Large, ruggedized, metal, 3.5 inch meter indicator.
- Uses 2 "D" cell batteries providing over 300 hrs of continuous use (alkaline)
CE marked.
- Sensitivity 0 - 10sv/hr range for monitoring barrier levels.
- Optional audible speaker with off-on switch for aural monitoring (ND-1000SA)
- Optional gamma remote probe with cable available (ND-1000SP)
Specifications
|
Ranges
|
ND1000S
|
0-1000, 0- 100, 0-10 Sv/hr
|
|
ND2000
|
0-10000, 0-1000, 0-100 Sv/hr
|
|
Efficiency
|
within 10% of full scale at standard temperature over operating range
|
|
Energy response
|
Gamma and X-ray from 50-1 500 KeV, ±20%
|
|
Saturation
|
In excess of 10 Sv/hr when calibrated and properly maintained
|
|
Environmental effects
|
Humidity limits: 0-99% non-condensing
|
|
Temperature limits
|
-30 degrees C to +50 degrees C (excluding batteries)
|
|
Detector
|
Halogen quenched, energy compensated geiger mueller tube effective length and diameter: 1.100 inch x 0.482 inch wall thickness: 30 mg/cm2
|
|
Controls
|
5 posn range switch: Off; X-100;X-10;X-1;and battery/high voltage test.
|
|
Calibration controls
|
Individual multiturn potentiometers for adjustment of each range
|
|
Calibration precision
|
Within 10% when calibrated with Cs137
|
|
Response time
|
90% of final reading within 10 seconds
|
|
Geotropisim
|
Within 2% of full scale in any orientation
|
|
Dimensions
|
4.4 inches wide x 7.75 inches long x 6.25 inches high (includes handle 11cm wide x 19.8cm long x 16cm high (includes handle)
|
|
Weight
|
3 pounds,13 ounces/1.73 kilograms (including batteries)
|
|
|
Direct Reading Dosimeters Meets ANSI Specifications N13.5 & N322
SPECIFICATIONS:
|
Detector:
|
Fiber electrometer mounted in an electrically conducting plastic ion chamber
|
|
Detector housing:
|
Very low permeability plastics - hermetically sealed
|
|
Rate response:
|
dose rate independent for gamma and x-radiation
|
|
M
|
138
|
0-200mr
|
725
|
0-5R
|
746
|
0-6OOR
|
|
138-S
|
0-2msv
|
730
|
0-20R
|
NEW
|
0-500mr
|
|
720
|
0-2R
|
742
|
0-200R
|
NEW
|
0-5msv
|
|
ODELS:
|
Operating temperature:
|
-20 deg c to +50 deg c
|
|
Relative humidity range:
|
up to 90%
|
|
Accuracy:
|
within ± 10% of true exposure
|
|
Warranty
|
2 year limited warranty from purchase date
|
|
Supplied with 1 year calibration certificate and calibration decal
|
|
Finish:
|
Natural matte black with metal clip
|
|
Compatibility: Compatible with dosimeter chargers that provide at least 195 volts, such as
- ND-100 and ND-120 (NDS PRODUCTS)
2000A (VICTOREEN) DCA 909 (DCA)
|
|
Dimensions:
|
Length: 4.5 inches; Diameter: 0.6 inches
|
|
Weight:
|
1.0 oz (25 grams).
|
|
|
|
|